Operating Messages; Error Codes; Instrument & Accessory Information - Orbeco SC450 Manual De Instrucciones

Ocultar thumbs Ver también para SC450:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Operating messages – Error codes
GB
Lo
btLo
Store
Date
Cal
RESULT
Time
Cal
E27 / E28 / E29
E 10 / E 11
E 20 / E 21
E23 / E24 / E25
E 22
E 70
E 71
E 72
E 73
SC450_V1_11.2010

Operating messages

Measuring range exceeded or excessive turbidity.
Result below the lowest limit of the measuring range.
Replace batteries, no further tests possible.
Battery capacity is too low for the display backlight;
measurement is still possible.
A user calibrated method is indicated by an arrow while the test result
is displayed (see "Factory calibration reset").

Error codes

Light absorption too great. Reasons: e.g. dirty optics.
Calibration factor "out of range"
Too much light reaching the detector.
Too much light reaching the detector.
Battery capacity was too low during measurement. Change battery.
CA:
Factory calibration incorrect / erased
CA:
User calibration incorrect / erased
A:
Factory calibration incorrect / erased
A:
User calibration incorrect / erased
Instrument & Accessory Information
IRiM
24 mm Sample Cells, 12 pack
Light shielding rings
Cleaning cloth for sample cells
Sample cell rack
Free Ammonia Reagent Set, 100 tests
Monochlor F Reagent, 100 pack
Free Ammonia Reagent, 4 ml
MC500-60
L197620
L197626
L197635
9999
RS5800
RP181-0
L531800
17

Publicidad

Tabla de contenido
loading

Tabla de contenido