Guidance and manufacturer´s declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environments listed below, and should
only beused in such environments:
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fl uctuations /
fl icker emissions IEC
61000-3-3
Guidance and manufacturer´s declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environments listed below, and should
only beused in such environments:
IEC 60601 –
Immunity test
test level
Electro-
± 6 kV
static discharge
contact
(ESD) IEC
± 8 kV air
61000-4-2
Power
frequency
(50/60 Hz)
3 A/m
magnetic fi eld
IEC 61000-4-8
Recommended separation distances between portable and mobile RF communication
The device is intended for use in an electromagnetic environment where radiated RF
disturbancesare under control. User can help prevent electromagnetic interference by keeping
the device at a minimum distance from portable and mobile RF communications equipment
(transmitters). Below table details the maximum output power of transmitter:
Rated
Seperation distance according to frequency of transmitter / m
maximum
output power
of transmitters
in Watt
150 kHz to 80 MHz
d = 1.2 √P
0.01
0.12
0.1
0.38
1
1.2
10
3.8
100
12
For transmitters rated at a maximum output power not listed above, the recommended separa-
tion distance d in metres (m) can be estimated using the equation applicable to the frequency
of the transmitter, where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
At 80 MHz and 800 MHz, the separation distance for the higher frequency
NOTE 1:
range applies.
NOTE 2:
These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and refl ection from structures, objects and people.
Compliance
Electromagnetic environment - guidance
RF energy is used only to maintain device's opera-
tion. Therefore, its RF emissions are so low that
Group 1
it's not likely to cause any interference in nearby
electronic equipment.
Class B
The device is suitable for use in all establishments,
Class A
including domesticestablishments, and those directly
connected to the public low-voltage power supply
networkthat supplies buildings used for domestic
complies
purposes.
Compliance
Electromagnetic environment - guidance
level
Floors should be wood, concrete or ceramic
± 6 kV contact
tile. If fl oors are covered with synthetic mate-
± 8 kV air
rial, the relative humidity should be at
least 30 %.
Power frequency magnetic fi elds should be at
3 A/m
levels characteristic of a typical location in a
typicalcommercial or hospital environment.
80 MHz to 800 MHz
d = 1.2 √P
0.12
0.38
1.2
3.8
12
33
800 MHz to 2.5 GHz
d = 2.3 √P
0.23
0.73
2.3
7.3
23