O P e r at I n g I n s t r u C t I O n s
_ Pour the contents of two containers of Kalk-X
liquid descaling agent (2 x 250 ml) into the boiler
and then fill the tank with warm water.
_ Allow the descaling solution to sit for at least
1 hour.
_ After this time, reopen the drainage screw and
empty the instrument.
gB
_ Accurately rinse the boiler tank twice using
clean water to remove any calcium residues.
_ The gasket on the safety cap
checked at this time, replace it if necessary.
_ After the descaling procedure, make sure that
the drainage screw has been closed correctly.
_ The Steamer X3 can now be refilled with water
and the safety cap screwed back on. Turn the
instrument on.
_ After the warm-up period, the instrument will
be ready for use.
safety aspects
keep the descaling agent out of the reach of
children! avoid contact with the eyes and skin!
In case of contact with the eyes, rinse with
plenty of water and seek medical advice. If
swallowed, seek medical advice immediately.
Do not use the descaling solution to clean the
external surfaces. Only use a detergent indicated
for aluminum/light metals to clean the housing.
rePlaCIng the saFety CaP gaskets
If the safety cap
and/or the gasket is damaged, the gaskets must
therefore be replaced.
14
should also be
^
no longer makes a good seal
^
PrOCedure FOr rePlaCIng the
saFety CaP gaskets
_ Insert a flat tool underneath gasket
_ Cut the gasket and remove.
_ Slip the gasket
over the threads and
BL
position it.
_ The gasket must be completed seated and
must not cover any of the screw threads.
technical data:
D/L/H:
275 x 265 x 455 mm
Weight:
12,5 kg
Mains supply:
230 V/50 Hz
Machine rating:
1700 W
E-fuse:
T10 A
Boiler volume:
3,7 l
Vapor pressure:
4,0 bar
Warm-up time:
approx. 20 min
accessories/Components:
Order No.:
116811
Kalk-X Liquid (4 x 250 ml)
116814
Small sieve Ø 50 mm
116815
Large sieve Ø 75 mm
116813
Funnel
516330
Clean Steamer
516340
Stymo Net
116911
Complete safety cap
116912
Safety cap gasket
1168391
Gasket kit for steam head
and lift.