Cleaning: home environment and hospital/clinic setting
After each use:
•
Pour away residual medication from the medication chamber.
•
Press the mouthpiece assembly release button to separate the mouthpiece assembly from
the handset.
•
Rinse the mouthpiece assembly thoroughly under running tap water.
•
Shake off excess water and allow to air dry fully before storing.
Daily cleaning
•
Wash the mouthpiece assembly by hand in a bowl of warm soapy water
(liquid dishwashing soap) for 2 minutes.
•
Rinse the mouthpiece assembly thoroughly under running tap water.
•
Shake off excess water and allow to air dry fully before storing.
Weekly disinfection
CAUTION: Risk of scalding. Use care around boiling water and in handling hot parts.
•
Prior to disinfection, ensure all parts are visibly
clean and free from dirt/debris.
Boil the mouthpiece assembly in water for
10 minutes. Ensure the medication chamber lid
is open and there is enough water in the pan to
prevent the mouthpiece assembly from touching
the bottom or boiling dry. Shake off excess
water and allow to air dry fully before storing.
Or
•
Immerse the mouthpiece assembly in a disinfectant of the Gluteraldehyde group
(testing performed with Korsolex Extra, 4% for 15 minutes). Rinse thoroughly according to
manufacturers instructions.
Mask cleaning
•
Wipe your handset clean with a clean damp cloth weekly or as needed. Do not use other
cleaning methods or solutions.
Handset cleaning
•
If using a mask and mask adapter, once a week: agitate for 2 minutes or soak for 10 minutes,
in warm soapy water (liquid dishwashing soap), rinse thoroughly under running tap water and
air dry fully.
•
After each cleaning procedure visually inspect for any remaining dirt/debris, if any remains
repeat the procedure. Visually inspect for any moisture before storing.
8
Maintenance
Mouthpiece assembly can be cleaned (boiled/disinfected) up to 52 times over 12 months.
To keep your InnoSpire Go working at the optimum level you should replace the mouthpiece
assembly and mask, if used, every 12 months as this is a consumable part.
LED Indication
LED Indication
SOLID GREEN
SOLID AMBER
FLASHES AMBER FIVE TIMES
FLASHES GREEN OR AMBER FIVE TIMES The treatment is finished
PULSES GREEN
ENGLISH
What it means
The battery is charged
There is enough charge for at least one
more treatment
There is not enough charge to take a treatment
The battery is being charged
9