2.
Intended Use
The testing device is designed for testing monopolar handles as well as monopolar
and bipolar handles and cables with a Valleylab, ERBE or Martin-compatible
connection.
3.
Operation
3.1
Switching on and functional test
Connect the adapter to the low-voltage socket with a 'push and turn' bayonet
fitting.
Plug the adapter into a locally provided socket; 100-240V AC possible.
Press the On / Off switch
A visible switch-on test is executed:
All the LEDs illuminate
The Ligator and the Insulation switch flash alternately,
the status indicator LEDs subsequently illuminate from left to right
1
Yellow –
Yellow –
The dummy resistance is activated in the circuit in order to test the sensor and
operational amplifier.
-
The switch-on test and the simulation test have been successfully completed
when the status indicators 2 and 4 illuminate orange.
The cable testing device is now ready for operation.
- or -
An error is indicated by the status indicator 4 flashing orange / red.
Please contact the Technical Service Center (see section "5.1 Technical
Service Center").
Caution:
Due to technical-measurement reasons, only one test item can be inserted at a
time!
Page 4
Testing Device
2
3
Green –
Blue –
–
Blue –
Red
Orange
-
4
Red
Green
Orange
MN030-423-IFU-M098-050230-S3-20151026