Testing Neutral Electrode Connecting Cables; Testing Bipolar Connecting Cables; Testing Ligator Connecting Cables - Bowa 050-230 Instrucciones De Uso

Comprobador para mangos de electrodos y cables
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 32
3.2

Testing neutral electrode connecting cables

Test the various neutral electrode cables (right-hand side of the cable testing device).
Select the appropriate plug-in mount for your cable; then connect the device side and
subsequently the instrument side.
Status indicator "4" illuminates green if the test item is ok. The test item should now
be moved back and forth to test whether there is a defective contact or not. The
Insulation switch starts to flash at the same time. This means that the respective
switch must be pressed to initiate the insulation test.
If the cable is ok, the status indicator "4" illuminates green; the test item should also
be moved back and forth. Subsequently remove the test item.
If the test item is not ok, the status indicator "4" illuminates red. In this case, remove
the test item and mark it accordingly.
3.3

Testing bipolar connecting cables

Select the appropriate plug-in mount for your cable. Then proceed as described for
the neutral electrode cables.
3.4

Testing Ligator connecting cables

Set the Ligator switch to "ON".
Select the appropriate plug-in mount for your cable; then connect the device side and
subsequently the instrument side.
The status indicator "4" illuminates green if the test item is ok. The test item should
now be moved back and forth to test whether there is a defective contact or not. The
insulation switch starts to flash at the same time. This means that the respective
switch must be pressed to initiate the insulation test.
If the cable is ok, the status indicator "4" illuminates green; the test item should also
be moved back and forth. Subsequently remove the test item.
If the test item is not ok, the status indicator "4" illuminates red or remains orange. In
this case, remove the test item and mark it accordingly.
MN030-423-IFU-M098-050230-S3-20151026