MANUAL DE INSTRUCCIONES CODIGO 80152 REV E
J
.P. SELECTA s.a.u.
DIC 2019
DELAYED START UP MODE:
1. Follow 1 item of MANUAL MODE.
2. Push (21) twice. Maintain it pushed and at the same time push
(22) or (23) to set delay time needed (between 1 and 24 hours).
During this operation, the indicator led TIME MODE (16) will
light.
3. If furthermore users need to set a run time, they have to follow
the instructions "Temporized time configuration".
4. Push START/STOP (24). The RUN indicator led (17) and
TIMER (18) will light. The remaining time will be able to see
on the display (19). Each decimal digit will express the tenth
part of an hour (6 minutes).
STARTING UP (T MODEL):
1. Check that the installation protocol has been correctly followed.
2. Check that the ventilation valve (3) is closed.
3. Load the material that you have to work. Never load more than
70% of the space in the chamber. Be sure that the hole in the
rear part of chamber is free.
4. Close the oven door.
5. Switch on the main switch (25). The display (19) will light.
6. Switch on the vacuum switch VACUUM PUMP (27).
7. When the vacuum level is the needed one (see vacuum gauge)
(4), disconnect the vacuum pump or the vacuum net with the
VACUUM PUMP switch (27).
NOTE: In this vacuum gauge you can read absolute pressure
so with the door the reading is 0.
8. Set temperature parameters (see "temperature configuration
parameters" of instruction STARTING UP TV MODEL).
9. When temperature process ends, it is recommended to pres-
surize the chamber with the control panel manual valve (3) in
order to avoid that the pressure induction put oil from vacuum
pump into the chamber.
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: www.grupo-selecta.com
(Sujetas a modificaciones sin previo aviso)
Tel 34 937 700 877 Fax 34 937 702 362
Pag.: 26