INSTRUCTION MANUAL CODE 80152 REV E
J
.P. SELECTA s.a.u.
DEC 2019
INSTALLATION
Put the oven on a flat, horizontal and level surface leaving a space
of 10 cm at the back and on each side of the equipment.
Connect the vacuum pump or vacuum net inlet to the outlet
placed at the rear part of the oven (30) and turn the manual valve
on (29) (the handle have to be positioned perpendicularly to the
equipment). Connect the plug at the socket which is on the rear
part of the oven (28). Turn the vacuum switch on.
If the work product can produce vapours, you should only use with
the oven, water jet pumps or hydrorotative pneumatic pumps. If an
air ballast pump is used, a water refrigerated condenser to deposit
gases and do not damage the pump has to be assembled between
the pump and the oven. Besides, a condensation collector has to
be added on each side of the pump. (suction and expulsion)
For maximum efficiency, do not place the oven in direct sunlight
or other sources of heat.
Connect the oven at the mains. Be sure that the voltage supply
is the same as the one indicated on the characteristics plate. Do
not use the equipment if it is NOT earthed.
If you have to change the plug cap, keep in mind the following:
Blue cable: Neutral.
Brown cable: Phase.
Yellow/green cable: Earth.
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: www.grupo-selecta.com
(They can be modified without notice)
Tel 34 937 700 877 Fax 34 937 702 362
Page: 21