IEC 60601-1-2: 2007 Table 2 Requirements
The Sensilift is intended for use in the electromagnetic environment
specified below. The customer or the user of the Sensilift should
assure that it is used in such an environment.
Immunity
IEC 60601
test
Test Level
Electrostat-
ic discharge
contact
(ESD) IEC
± 8 kV air
61000-4-2
Electrical
± 2 kV for
fast
transient/
supply lines
burst IEC
61000-4-4
± 1 kV line(s)
to line(s)
Surge IEC
± 2 kV line(s)
61000-4-5
to earth
<5 % UT
Voltage
(>95 % dip
dips, short
in UT) for
interruptions
0,5 cycle
and voltage
40 % UT (60
variations
% dip in UT)
on power
for 5 cycles
supply input
70 % UT (30
lines
% dip in UT)
IEC
61000-4-11
<5 % UT
(>95 % dip
in UT) for 5 s
40
Compliance
± 6 kV
contact
± 8 kV air
± 2 kV for
power
power
supply lines
± 1 kV line(s)
to line(s)
± 2 kV line(s)
to earth
<5 % UT
(>95 % dip
in UT) for
0,5 cycle
40 % UT (60
% dip in UT)
for 5 cycles
70 % UT (30
% dip in UT)
for 25
cycles
<5 % UT
(>95 % dip
in UT) for 5 s
level
environment – guidance
Floors should be wood,
concrete or ceramic tile. If
± 6 kV
floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment. If
the user of the Sensilift
requires continued
operation during power
mains interruptions, it is
for 25
recommended that the
cycles
Sensilift be powered from
an uninterruptible power
supply or a battery.
Electromagnetic