V- Positioning the tubes

VI- Malfunctions list:
Problem
The pump is not working
properly
Commands are not
working
Débit de l'irrigation trop
bas
Débit de l'aspiration trop
bas
Tube lifetime too low
Beats into the tube
Abnormal noise coming
out of the pump
NT001
O – 2015/11
Incorrect
Figure 2 How to position the tubes: front and top views
Possible reason
No electrical power
The pump wheel stalls
No control of the irrigation
No control of the suction
Bad setting of the flow rate
Tube partially or totally
clogged
Wearing of the tube
Tube partially or totally
clogged
Wearing of the tube
Tube/pump incompatibility
Tube/pump incompatibility
Bad setting up of the tube
into the pump block
Use generating important
beats, due to the pump,
the speed of the pump,
the pressure in the tube
Deterioration of the device
Correct
Solution
Verify that power supply switch is on
Verify the power cord connection
Verify the tube connection
Verify that the pneumatic petal is
switched on
Verify that the handle is switched on
Régler le débit à l'aide du
potentiomètre
Installer une nouvelle tubulure stérile,
référence Integra MicroFrance
CM217S
Set up a new sterile tube, Integra
MicroFrance reference CM217S
Set up a new sterile tube, Integra
MicroFrance reference CM217S
Set up a new sterile tube, Integra
MicroFrance reference CM217S
Set up a new sterile tube, Integra
MicroFrance reference CM217S
Set up a new sterile tube, Integra
MicroFrance reference CM217S
Set up a new sterile tube, Integra
MicroFrance reference CM217S
Contact Integra MicroFrance
Contact Integra MicroFrance
Page 38 sur 91