XVI- Electromagnetic emissions and immunity
Origin of
Frequency
interference
range
0.15 to 0.5 MHz
AC supply
0.5 to 30 MHz
30 to 230 MHz
Casing
230 MHz to
1 GHz
Harmonic
0 kHz to 2 kHz
currents
Flicker
-
Origin of
interference
Electromagnetic
field with amplitude
modulation radio
frequency
Electrostatic
discharges
Magnetic field in
power grid
frequency
NT001
O – 2015/11
Limits
Quasi-peak:
79 dBµV
Average
66 dBµV
Quasi-peak
73 dBµV
Average
60 dBµV
30 dBµV/m
at 10 m
37 dBµV/m
at 10 m
Class A
Pst, Plt
Dc, D(500),
DMAX
Casing access
Test specifications
80 to 2500 MHz
3 V/m
AM 80%
1 kHz
+/- 6 kV in contact
+/- 8 kV in air
50 Hz
3 A/m efficient
Emissions
Basic standard
CISPR 11
Class A
Group 1
CISPR 11
Class B
CEI 61000-3-2
CEI 61000-3-3
Immunity
Suitability
Basic standard
criteria
See
CEI 61000-4-3
standard
See
CEI 61000-4-2
standard
See
CEI 61000-4-8
standard
Results
The statistical
Satisfactory
evaluation of the
standard is applicable.
The statistical
Satisfactory
evaluation of the
standard is applicable.
Satisfactory
Nothing to report.
Considering the low
power consumption,
Satisfactory
the pump satisfies the
limits without additional
measurements.
Results
Satisfactory
Satisfactory
Satisfactory
Comments
Comments
The specified level
is that of the
carrier, with
efficacy value,
before modulation.
See the basic
standard for the
applicability of
discharges in
contact and/or in
the air.
-
Page 44 sur 91