Preparation for cleaning/disinfection

Directly after each treatment, all parts of the INQUA
be cleaned of medication residue and contamination.
In order to help reduce water pollution, do not dispose of the medica-
tion residue in the waste water system. Wipe it up with tissues and dis-
pose of the tissues with non-recyclable waste.

Cleaning with warm drinking water

Clean all parts of the nebuliser thoroughly for 5 minutes with
warm drinking water (approx. 40 °C) and a little dishwashing
liquid with no skin protecting or disinfecting additives (dosage
according to the manufacturer's recommendations) until all resi-
dues of the inhalation solution and other contaminants have been
removed.
If necessary, use a small brush to remove dirt. For hygiene rea-
sons, this brush must then be used exclusively for cleaning the
nebuliser.
If this has not been done, disconnect
the connection tubing from the
®
INQUA
nebuliser.
Dismantle the nebuliser into its individ-
ual parts.
®
nebuliser must
en
45