• Screw the front section on fully.
• Screw on the piston button.
• Pull the piston back until you hear a locking "click".
• Replace the front canula.
Before use, ensure that the product is
assembled and the piston is working
correctly. The front canula should be set
into its housing. The piston should be in
the rear position.
• Connect the AspeO® to the suction
system.
• Keep the front canula close to the
implant site while drilling.
TO DISPOSE OF THE BONE FRAGMENTS:
• Remove the front canula.
• Engage the piston by pushing it forward
as on a syringe.
Collected bone fragments may then be
deposited in a cup or directly placed on
the site to be filled.
TO
RESUME
HARVESTING:
• Pull the piston back to its initial rear
position. A "click" will confirm its correct
positioning.
• Replace the front canula.
• Reconnect the AspeO® to the suction system.
VI. SURGERY PROCEDURE
BONE
FRAGMENT
15