MANUAL DE INSTRUCCIONES
Fig. 8.5.5 Setting alarm margins
Fig. 8.5.6 Password entry
Fig. 8.5.6.1 Temperature Offser modification
Fig. 8.5.7 Password modification
Fig. 8.5.8 Logs download
J
.P. SELECTA s.a.u.
CODIGO 80390 REV A
8.5.5 Alarms
When entering this setting, screen Fig 8.5.5 appears. Use the Up/Down
arrows to select the value of over temperature or low temperature you
want for the temperature range.
This setting is simply a warning, it does not stop the operating cycle. This
notice will be reflected in the memory log.
Press the BACK key to return to the settings menu or the HOME key to
return to the initial screen.
8.5.6 Temperature Offset Adjustment
The Offset adjustment is protected by a password to prevent any user
from modifying this parameter.
Press the CAL button to access this setting.
If the CAL button is pressed without entering the password, it will indica-
te that a valid password value is necessary to continue with the action.
Therefore, it is necessary to select ENTER PASSWORD option before
pressing the CAL button.
The default password value for the calibrator is "0000".
Press on the temperature Offset parameter and use the Up/Down keys
to modify this value.
Press the BACK key to return to the settings menu or the HOME key to
return to the initial screen.
The password value remains valid until the oven is switched off and on
again. Perform this action to prevent another user from modifying the value.
8.5.7 Password Change
If you want to change the factory password, use this menu and follow the
instructions that appear to change the password.
Press the BACK key to return to the settings menu or the HOME key to
return to the initial screen.
8.5.8 Logs Download
To download the data stored in the control, select the option LOGS
DOWNLOAD. Screen Fig. 8.5.8 will appear.
Connect a memory pen drive to the front connector (4) and follow the
instructions on the screen.
There are two download options: either download all logs or the last 10
only.
During the loading of data in the USB memory, information about % of
loading appears.
Press the BACK key to return to the settings menu or the HOME key to
return to the initial screen.
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - http://www.grupo-selecta.com
01/2021
Pag.: 27
Tel 34 93 770 08 77 Fax 34 93 770 23 62

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

3000982

Tabla de contenido