Specifications - VWR 444-7084 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Euro plug: 444-7084
UK plug: 444-7085
S
PECIFICATIONS
Overall dimensions (L x W x H):
Interior dimensions (L x W x H):
Tray dimensions (L x W):
Electrical (50/60 Hz):
Fuses:
Temperature range:
Temperature uniformity:
Speed range:
Speed accuracy:
Timer:
Orbit:
Maximum weight capacity:
Controls:
Tray material:
Ship weight:
5
Swiss plug: 444-7086
64.8 x 35.6 x 40.6cm
34 x 30.5 x 24.1cm
33 x 27.9cm
230 volts, 5 amps, 450 watts
5mm x 20mm, 5 amp quick acting
ambient +5°C to 65°C
±0.5°C at 37°C
15 to 500rpm
±1%
1 second to 9999 minutes
(increased in 1 second increments)
19mm
15.9kg
see page 4
aluminum
37.7kg
O
I
PERATING
NSTRUCTIONS
The Incubating Orbital Shaker has been designed for the temperature,
speed and time functions to work independently of one another. The
temperature and speed can be reset without resetting the timer and the
timer can be stopped and started without interrupting the heating and
shaking functions.
1. Getting ready:
a. Plug the power cord into a properly grounded outlet. The standby
indicator light will illuminate, verifying power to the unit.
b. Press the standby button to move the unit from standby mode.
The standby indicator light will turn off and the temperature,
speed and time displays will illuminate, displaying the previously
used settings.
2. Setting temperature:
a. Press the up/down arrows below the temperature display until you
reach the desired temperature. When you release the button, the
display will blink off and then on indicating the new set temperature
has been accepted.
b. Press the on/off button to start the heating function. The indicator
light below the temperature display will illuminate to indicate the
heating function is in use and remain lit until heating has ceased.
c. Temperature adjustments can be made without interrupting heating
by using the up/down arrows below the temperature display. After
the change has been made and you release the button, the display
will blink off and then on indicating the new set temperature has
been accepted.
d. To stop the heating function, press the on/off button below the
temperature display. The heat indicator light will turn off.

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

44-7085444-7086

Tabla de contenido