Control solution testing
Running a control test lets you know that your meter and test strips
are working properly to give reliable results. You should run a con-
trol test when:
• You use the smartLAB
• You open a new vial of test strips.
• You think the meter or test strips might be working incorrectly.
• You drop the meter.
• You have repeated a test and the test results are still lower or
higher than expected.
• You are practicing the test procedure.
Note: Professional users are instructed to follow federal, state, and
local guidelines concerning QC practices.
About the smartLAB
• smartLAB
®
• smartLAB
®
tion or injection.
• use only with smartLAB
• Shake the smartLAB
• Write the date you first opened the bottle on the bottle label.
• The smartLAB
the date the vial is opened or until the „Expiration Date" on the
bottle, whichever comes first.
• Do not use smartLAB
Date
®
mini blood glucose meter for the first time.
control solution
®
control solution is for in vitro diagnostic use only.
control solution is not intended for human consump-
®
®
control solution well before each use.
®
control solution is durable for three months from
control solution that is past the Expiration
®
pro test strips.
23