4 CALIBRATION

Before calibrating for the first time:
Shake the probe gently to remove any air bubbles from underneath the glass bulb.
Remove the storage cap and rinse the probe with demineralised water.
4.1 pH
Refer to the pH transmitter's instructions to select the correct calibration type.
Rinse the probe with the freshly prepared first buffer solution (pH 6.88 or 7.00), ideally at
the temperature of the sample to be measured.
Immerse the probe (to a depth of 4 to 5cm) and wait for stable measurements before
calibrating.
Rinse the probe, first with demineralised water and then with the second buffer solution
(pH close to that of the process, 4.00 or 9.22).
Immerse the probe (to a depth of 4 to 5cm) and wait for stable measurements before
calibrating.
Rinse the probe in demineralised water.

4.2 REDOX

Rinse the probe using a standard, freshly prepared quinhydrone solution, ideally at the
temperature of the sample to be measured.
Immerse the probe (to a depth of 4 to 5cm) and wait for stable measurements before
calibrating.
The potential is measured with respect to the Ag/AgCl reference system. Add 200mV for an
E
reference potential.
H2

5 CLEANING AND RECONDITIONING

These electrodes are considered as consumable items and therefore have a limited lifespan,
largely dependant on the user's application. Under normal conditions, a typical lifespan would be
about one year.
Before replacing them, however, simply cleaning the electrodes may restore them.
Such cleaning will also be necessary whenever the following are observed:
An offset from zero or a measurement drift (the reference system is contaminated).
A significant drop in response time and a loss of slope (there is fouling on the glass diaphragm
or the platinum).
Provided the glass diaphragm is neither scratched nor cracked, follow one of the cleaning
procedures below:
Salt deposits: HCl 0.1M (5 mins) + NaOH 0.1M (5 mins)
Grease or oil film: Detergent + water
Reference junction fouled: KCl 3M @ 60-80 °C for 10 mins, then cool in KCl 3M@25 °C
Calcium, metal hydroxide or protein deposits: HCl 10% for 10 mins
Redox measurements:
Deposits/ wear: aqua regia (1/1 concentrated HCl / concentrated HNO
only be carried out by qualified individuals.
If the electrode's response is not improved, replace it.
); this process must
3
5/6