Configuration; Technical Specifications - RS Pro RS-8806S Guía De Inicio Rápida

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 19
Non-Contact Forehead IR Thermometer Quick Start Guide

4. Configuration

1-IR Sensor
2-LCD Display
3-Mode Selection
4-Down Button
2
5-Up Button
6-Mode Button
7-Measurement Trigger
8-Battery Cover
5. Indicator
1-Surface mode Symbol
1
2-Body mode Symbol
2
3-Digital readout
4-Battery Symbol
3
5-The order number
6-Save data readout
5
7-Temperature °C(Celsius)/
°F(Fahrenheit) Scale
8-Buzzer symbol
2
3
1
4
5
6
7
8
8
7
4
6
Non-Contact Forehead IR Thermometer Quick Start Guide

6.Technical Specifications

Normal Conditions of Use
Display Resolution
0.1°C (0.1°F)
Operating Temperature
10 to 40°C (50 to 104°F)
Storage Temperature
0 to 50°C (32 to 122°F)
Humidity Rate
≤85%
Power
DC 3V (2 x "AA" batteries)
Size
128 x 74x 36 mm / 5 x 2.9 x 1.4 in (L x W x H)
Weight
Gross 125.4g / Net 104.5g
Measuring Range
In Body Mode
32.0 to 42.5°C (90 to 108°F)
In Surface Temp Mod
0 to 60°C (32 to 140°F)
Accuracy
±0.3°C (0.54°F)
Emissivity
Fixed at 0.99
Measuring Distance
1 cm – 10 cm (0.39 in – 3.9 in)
Automatic Stop
7 sec.
Non-contact Body Infrared Thermometer Precision
32 to 35.9°C / 93.2 to 96.6°F
36 to 39°C / 96.8 to 102.2°F
39 to 42.5°C / 102.2 to 108.5°F
±0.3°C / 0.5°F
According to ASTM
±0.2°C / 0.4°F
Standard E1965-1998
(2003)
±0.3°C / 0.5°F
3

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

2010210

Tabla de contenido