Non-Contact Forehead IR Thermometer Quick Start Guide
9. Measuring Operation
9-1.Choosing The Temperature Unit – F1 Function
Press "MODE" button for 2 seconds, the screen displays: F1. Select "DOWN" for degrees
Celsius, "UP" for degrees Fahrenheit.
9-2.Alarme Setup – F2 Menu
Press "MODE" touch for 2 seconds, the screen displays: F1.Press twice "MODE" touch to
get F2. Select "UP" to increase the threshold by 0.1°C (0.1°F), "DOWN" to reduce it by 0.1°C
(0.1°F).
Note: The alarm threshold default value is 38°C (100.4°F).
9-3.Total Difference – F3 Menu
To adjust the total variation of your the Non-contact Body Infrared Thermometer.
Press "MODE" button for 2 seconds, the screen displays: F1. Press "MODE" button two
times to get F3. Select "UP" to increase the difference by 0.1°C (0.1°F), "DOWN" to reduce
itby 0.1°C (0.1°F).In the cases of seasonal or environmental changes a verification and
adjustment should be carried out.
Note: This function is only effective Body
9-4.Buzzer ON/OFF – F4(F3) Menu
Press "MODE" button for 2 seconds, the screen displays: F1. Press three(two) times "MODE"
button to get F4(F3).
Select "UP" to open the buzzer (a sound icon " " is displayed on the LCD screen), press
"DOWN" to stop it (the icon disappear).
9-5.Exiting The Setting Mode
Press "MODE" button until the screen turns off.
9-6.The Non-contact Body Infrared Thermometer is specially designed to take the body
temperature of a human being. For this, use the Body mode. Measurement range for Body
mode: 32 to 42.5°C (86 to 108°F)
You can also use the Non-contact Body Infrared Thermometer to measure the temperature of
an area or an object, a food, a liquid or a room temperature. For this, use the Surface mode.
Measurement range for Surface mode: 0 to 60°C (32 to 140°F)
Important: The area temperature differs from the internal body temperature. To obtain the
internal temperature always use the Body mode.
Please make sure to select the Body mode for an internal temperature reading and the
Surface mode for an external area reading (bottle, bath, room...).
6
Non-Contact Forehead IR Thermometer Quick Start Guide
9-7.Data Memory
Data memory automatically after temperature measurements, which will display at the right
corner of LCD. Press "UP" or "DOWN" button to display the last temperature measurement.
Under the condition of power off, press "UP" and "DOWN" button together for two seconds,
to display the last temperature measurement.
At this mode if change the order number to 0 and press "MODE" button, then delete all
memory data.
9-8.Changing The Batteries
Display: when the LCD screen displays " ", the battery is used.
Operation: Open the lid and change the batteries, taking great care with the correct
positioning. A mistake with this could cause damage to the apparatus and compromise the
guarantee of your Non-contact Body Infrared Thermometer. Never use rechargeable batteries.
Use only batteries for single usage.
Remove the battery from the instrument if it is not required for extended periods of time in
order to avoid damage to the thermometer resulting from a leaking battery.
9-9.Longevity Use
The Non-contact Body Infrared Thermometer was conceived for an intense and professional
use, its longevity is guaranteed for 40000 readings.
10.Advice
•The protective glass over the lens is the most important and fragile part of the thermometer,
please take great care of it.
•Do not recharge non rechargeable batteries, do not throw in fire.
•Do not expose the thermometer to sunlight or water.
11.Maintenance and Cleaning
• The Infrared Sensor is the most precise part, must be protected carefully.
• Clean the device with a cotton bud lightly moistened with 70% alcohol.
• Do not clean the device with corrosive detergent.
• Keep the device away from water or other liquid.
• Store the device in a dry environment, and keep it away from dust and direct sunlight.
7