OHAUS CKW Serie Instrucciones De Instalación Y Operación página 9

Juego opcional de batería para el indicador y para las básculas
Ocultar thumbs Ver también para CKW Serie:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

CKW Battery Option Kit
Operating Under Battery Power (Cont.)
Conditions that Increase Battery Run Time
Use of Sleep Mode with a number other then 0
Digit and/or Graph display set to lower brightness level
Setting Graph Display to Segment Mode
Use of Auto-Off Mode with a number other then 0
Minimal use of Serial communication
Remove loads on the pan to reduce Bar Graph use
Minimal use of Beeper signal
Operation at nominal room temperature
Visible Differences
There are minor differences seen when operating only from battery power.
1. At power on, the display will include the message "On bAt".
2. The initial power on display will cycle through each display section
individually rather then turning on all at one time.
3. The Bar Graph menu will show fewer segments during the brightness level
adjustment.
4. The Sleep mode is indicated with a flashing decimal point rather then by a
continuous decimal point.
Extending Battery Run Time with Sleep and Auto-Off Modes
Sleep Mode:
Because the display usually consumes most of the power from the system,
switching off the display when not required will significantly increase the run time
of the batteries. The Sleep mode temporarily turns off most of the display
segments when there is a lack of activity for a specified number of minutes. The
time selected for the Sleep mode may be consistent with the maximum time a user
needs to view new weighing results on the display while not physically interacting
with the system. The minimum setting is 1 minute.
Auto-Off Mode:
Use of the Auto-Off mode will conserve the most battery power. However, unlike
the Sleep mode, use of the Auto-Off requires that the system be powered back on
each time.
Conditions that Decrease Battery Run Time
Sleep Mode set to 0
Digit and/or Graph display set at high brightness level
Setting Graph Display to Bar or Block Mode
Auto-Off Mode set to 0
High use of Serial communication
Loads left on the pan with a bar Graph display
Excessive use of Beeper signal
Operation at elevated or very low temperatures