Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 31
TC-Kit (+)
Expansion module 230/115 V
PP.2 N
PP.2 L
Blank PP.2
PP.1 N
PP.1 L
Blank PP.1
Status n.o.
Status com.
Status n.c.
GP +24V
GP -
4.5.1.1 115 V and 230 V switching power supply
Connect the power supply for the sample gas cooler to the
switching power supply. The earth conductor must be connec-
ted to the earth bolt on the housing. Please note the polarity
when connecting.
4.5.1.2 24 V DC
The voltage supply for the sample gas cooler must be connec-
ted to terminals 8 and 9 on the power board. The earth con-
ductor must be connected to the earth bolt on the housing.
The operating voltage of the pumps which may optionally be
connected is 24 V DC. Please note the polarity when connect-
ing.
BX440028 ◦ 10/2020
Ter-
230 V 115 V
Note
minal
1
Signal relay 1,
Connec-
white
tion ex-
pansion
2
Signal relay 2,
and con-
orange
N
L
troller
3
+24 V, red
L
N
board
4
GND, blue
5
blue
Sample
gas
N
6
red
L
pump
(GP 24 V)
7
Status n.c.,
Status
red
output
8
Status com.,
yellow
9
Status n.o.,
white
10
white yellow
Con-
densate
11
yellow white
pumps
12
blue
blue
(230/115
13
white yellow
14
yellow white
15
blue
blue
16
Functional
Power
earth
supply
17
Lead, black
18
N-lead, blue
19
Lead, black
Connec-
tion SNT
20
N-lead, blue
and ex-
pansion
board
21
-
22
-
23
Lead, brown self-reg-
ulating
24
N-lead, white
heated
line
25
N-lead, blue
regu-
lated
26
Lead, black
heated
line

4.6 Signal outputs

Plug connection (optional)
The unit is optionally equipped with a EN 175301-803 plug for
the status output.
Plug numbering
The clamping area has a diameter of 8 – 10 mm (0.31 –
 0.39 inch).
Connection regulated heated line (optional)
Connection
V)
Connection to board see chapter Wiring diagrams.
Connecting other options
Optional
Condensate pump
Gas pump
When using options by other manufacturers, use suitable
-
fuses.
Alarm contact (S2)
Analog output (S3)
Pin
Assign-
Line type
ment
1
L 230/115 V
regulated heated line
2 N 230/115 V
3
N 230/115 V self-regulating heated line
4
L 230/115 V
5
Pt100
regulated heated line
6
Pt100
7
PE
Supply voltage
max. power input
230 V/115 V
0.025/0.044 A
24 V
0.8 A
Bühler Technologies GmbH
1 - N/C
2 - N/C
3 - GND
4 - 4-20 mA out
5 - shield
15