Section 1 Safety

Never operate the bath with the immersion circulator removed. 
Do not mount the immersion circulator backwards; the line cord
could contact the reservoir fluid. Ensure the electrical cords do not
contact any of the plumbing connections or tubing. 
Never place the circulator in a location or atmosphere where
excessive heat, moisture, or corrosive materials are present. 
Ensure the tubing you select meets your maximum temperature and
pressure requirements. 
Ensure all communication and electrical connections are made prior
to starting the circulator. 
Never operate the circulator without fluid in the reservoir. 
Oil-based fluids expand when heated. Fill the reservoir only to the
minimum level. 
Use only the approved fluids listed in this manual. Using other fluids
voids the warranty. 
Other than water, before using any approved fluid, or when
performing maintenance where contact with the fluid is likely, refer to
the SDS and EC Safety Data Sheet for handling precautions. 
When using ethylene glycol and water, check the fluid concentration
and pH on a regular basis. Changes in concentration and pH can
impact system performance. 
Do not clean the circulator with solvents, use a soft cloth and water. 
Drain the reservoir before the bath is transported, and/or stored. 
Always turn the circulator off and disconnect the supply voltage from
its power source before moving or before performing any service or
maintenance procedures. 
Transport the circulator with care. Sudden jolts or drops can damage
its components. 
Refer service and repairs to a qualified technician. 
1-9