Calibration Menu - TechGrow T-1 PRO Serie Manual Del Usuario

Ocultar thumbs Ver también para T-1 PRO Serie:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Calibration
The TechGrow T-1 Pro comes with an unique
calibration feature that allows you to calibrate the
device in case of large deviation.
Only calibrate the sensor if you suspect a
deviation.
Experience has shown that the SenseAir®
sensors we use function without any problems for
at least two years without calibration being
necessary.
The sensors are calibrated with the highest
precision during manufacture, however it could
occur after a year in use that the sensor shows
a deviation and needs to be calibrated.
Preparation for calibration:
1. Place the CO2 sensor in a CO2 neutral
environment (+/- 400ppm), such as a balcony
or your garden for at least half an hour, or in a
sealed and well-ventilated room for at least 8
hours. This is best done during daytime, because
the CO2 values will be closest to 400 ppm.
During nighttime, plants and trees emit CO2 and
the outdoor CO2 values will be closer to 450 ppm.
If the minimal value is not much lower than 370
ppm or much higher than 430 ppm, it is not
necessary to calibrate the sensor. That is because
the value of neutral/clean outdoor air is between
380 ppm and 420 ppm. A deviation of +/- 5% is
common with this type of measurement
equipment. Please keep this in mind when setting
your CO2 values. In case of a greater deviation,
calibration is necessary.
Please continue with step 2.
2. To calibrate the sensor, you have to switch DIP
switch 1 (see image) from position 0 to
position 1 (ON). This is to prevent any
unintentional calibration caused by any
electronic/dimmable control gear, which can
cause much disturbance to our measurement
equipment. This manual security has been built in
for exactly that reason.
3. Open the sensor case (see image A)
4. Locate DIP switch 1 (see image B).
5. Switch DIP switch 1 to position 1 (ON) using a
pen or a small screwdriver (image C).
6. The sensor can now be calibrated.

Calibration menu:

1. Disconnect the power supply to the T-1 Pro
CO2-controller.
2. Re-connect the power after 20 seconds.
3. The display will fi rst show the type
and software version of the controller.
TechGrow
T-1
Pro-Series
Software v. 1.X
4. After 5 seconds the question appears to
continue calibrating. Choose Yes.
CO2 calibration
Continue?
Yes
=> No
5. If you do nothing, or choose No, the controller
returns to the main menu without calibration.
Foto A
Foto B
Foto C

Publicidad

Tabla de contenido
loading

Tabla de contenido