7.
To Stop the bleaching device at any time press the START / STOP button. If the device is not
manually stopped it will stop after the selected time has elapsed.
Note that the fan will continue to work for some time
in order to cool down the unit.
8.
At the end of the day turn off the device by holding down
the Start / Stop button for 3-4 seconds
or by disconnecting it from the power network.
9. In case of overheating the thermal protection is triggered
and the device will cease to emit light. The display will indi-
cate "Oh" and the fans will turn on for 1 minute. After cooling,
the device will continue its normal operation, but stays in
Pause mode. Overheating can only happen in an emer-
gency situation or by accidental clogging of the cooling
vents. For this reason the cooling vents located on the bot-
tom of the bleaching head must always be unobstructed in
order to allow air circulation.
VII. DAILY CARE AND MAINTENANCE
1.
Disinfection of the PVC window of the bleaching head
For each patient, the emitting window is cleaned with
a cotton swab soaked in an alcohol solution.
2.
Cleaning the device:
To disinfect the device and its parts, spray disinfectant on a piece
of soft cloth or cotton and clean.
Do not use abrasives or solvents,
as this may damage parts of the device!