Guidance and Manufacturer's Declaration – Electromagnetic
Immunity
The Micro CO is intended for use in the electromagnetic environment
specified below. The customer or the user of the Micro CO should assure
that it is used in such an environment.
Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
IEC 60601 test
level
Contact: ± 8 kV
Air:
±2 kV, ± 4 kV, ±
8 kV, ± 15 kV
± 2 kV
100 kHz
repetition
frequency for
power supply
lines
± 0,5 kV, ± 1 kV,
± 2 kV
0 % U
; 0,5
T
cycle
At 0°, 45°, 90°,
135°, 180°,
225°, 270° and
315°
0 % U
; 1 cycle
T
and
70 % U
; 25/30
T
cycles
Single phase:
at 0°
and
0 % U
; 250/300
T
cycle
Compliance
Level
Contact: ± 8 kV
Air:
±2 kV, ± 4 kV, ±
8 kV, ± 15 kV
not applicable,
(battery powered)
not applicable,
(battery powered)
not applicable,
(battery powered)
33
Electromagnetic
Environment
-Guidance
Floors should be
wood, concrete or
ceramic tile. If
floors are covered
with synthetic
material, the
relative humidity
should be at least
30%.
Mains power
quality should be
that of a typical
commercial
or hospital
environment.
Mains power
quality should be
that of a typical
commercial
or hospital
environment.
Mains power
quality should be
that of a typical
commercial
or hospital
environment. If the
user of the Micro
CO requires
continued operation
during power main
interruptions, it
is recommended
that the Micro CO
be powered from an
uninterruptible
power supply or a
battery.