Vyaire MicroCO Manual De Funcionamiento página 59

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 145
connections to the gas cylinder are secure, and that the gas cylinder is
not empty, and then repeat the calibration procedure.
If the above message is repeated, the CO meter must be returned to
Vyaire, or an authorised agent, for sensor replacement. Sensor life is
greater than 2 years. The sensor life depends upon both the amount
of exposure to CO and other gases, particularly solvents such as
alcohol and cleaning fluids. If the signal from the CO sensor is too
high, a new calibration value will not be stored and the following will
displayed when calibration is attempted:
The most likely cause for this is using an incorrect concentration of
calibration gas. Ensure that the concentration of calibration gas is
correct (20ppm) that the connections to the gas cylinder are secure,
and then repeat the calibration procedure.
In order to stop any gas leak from the can after calibration, ensure
that the control knob is tightened firmly.
Important Notes:
Only certified calibration gas from a reputable source should
be used.
Ensure that no CO is present on the sensor for 3 minutes
before starting the calibration procedure.
Ensure that the instrument and gas cylinder have stabilized at
room temperature before calibrating.
57

Publicidad

Tabla de contenido

Solución de problemas

loading

Productos relacionados para Vyaire MicroCO

Tabla de contenido