Servicing; Trouble Shooting Information - CareFusion MicroCo Manual De Funcionamiento

Tabla de contenido

Publicidad

Servicing

If your unit requires service or repair please see page 73 for contact
details. A full service manual including circuit diagram and parts list is
available upon request.

Trouble Shooting Information

Should you encounter problems operating your MicroCO meter, please
consult the table below:
Problem
MicroCO cannot be
switched on or "bat"
is displayed
Battery life is
shorter than
expected
"CEL" is displayed
"Err" is displayed
"bt2" is displayed
"gAS" is displayed
Possible Cause
Batteries are flat
Unit not being
switched off
Fuel cell is
depleted
Gas cylinder empty
Internal battery
has expired
Fuel cell is
depleted
Fuel cell is
contaminated by
residual gas or
solvents
61
Solution
Change the battery
Turn the unit off after
use
Return unit for cell
replacement
Check valve on cylinder
for contents and replace
cylinder if necessary
Check the calibration
value on the cylinder is
20ppm.
Return unit for battery
replacement
Return unit for cell
replacement
Allow unit to be
exposed to clean air.
Turn unit ON and await
message to disappear,
if not return unit for cell
replacement.

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido