Cleaning and Disinfecting
Warnings:
- Inspect the product for damage and replace if necessary.
- DO NOT allow water or cleaning solution to enter the product. Make sure the inflation valves are closed.
- DO NOT machine wash or sanitize or machine dry overlay products.
- Cleaning and disinfecting are separate processes. Cleaning must precede disinfection. Before use by a different individual: clean, disinfect, and check the
product for proper functioning.
- Clean the product regularly and when it becomes soiled.
- Follow all manufacturer bleach container and germicidal disinfectant safety guidelines.
- Note about sterilization: High temperatures accelerate aging and will damage the product assembly. The products featured in this manual are not packaged
in a sterile condition, nor are they intended or required to be sterilized prior to use. If institution protocol demands sterilization: First, follow the cleaning and
disinfecting instructions; then, open the inflation valves and use the lowest possible sterilizing temperature, but not to exceed 70°C (158°F), for the shortest
time possible. DO NOT autoclave.
Cautions:
-
Use of the following may cause damage to the product: abrasives (steel wool, scouring pads); caustic, automatic dishwasher detergents; cleaning products
that contain petroleum or organic solvents, including acetone, toluene, Methyl Ethyl Ketone (MEK), naphtha, dry cleaning fluid, adhesive remover; oil-based
lotions, lanolin, ozone gas; ultraviolet light cleaning methods.
-
DO NOT expose to direct sunlight.
-
Rinse thoroughly. Detergent residue may cause air cells to stick together.
Recommendation: When washing by hand, use room-temperature water.
Note: Deflated air cells may be easier to thoroughly clean. To deflate the air cells: with the inflation valve(s) open, roll
up and/or compress the cushion, and then completely close the inflation valve(s).
To Clean Overlay Section(s): Remove all linens and straps. Close valves. Unsnap overlay sections. Place each
overlay section in a large sink. Hand wash, using mild liquid hand soap, hand dishwashing liquid, laundry detergent, or
a multipurpose detergent (follow the product label instructions). Gently scrub all product surfaces, using a soft plastic
bristle brush, a sponge or a cloth. Rinse with clean water. Air dry.
To Disinfect the Overlay: Follow the cleaning instructions, using 1 part household liquid bleach per 9 parts water.
Keep the product wet with the bleach solution for 10 minutes. Rinse thoroughly with clean water. Air dry.
Care Symbols
Hand wash.
Do not tumble dry.
ROHO PRODIGY Mattress Overlay Operation Manual
Line dry in the shade.
Do not machine wash.
6
+
Bleach (1 part bleach: 9 parts water).