Operation/Alarm Log; Setting Log Interval - Phcbi MCO-230AIC Instrucciones De Operación

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

OPERATION/ALARM LOG

Setting log interval

The incubator can save operation log data (chamber temperature, CO
outer door). Use the following procedure to set the log interval (interval of acquiring the operation log).
1. Go to the Log screen (Menu
"Setting" to display the Setting screen.
2. Enter the log interval. Press "Apply" to save the
entered value.
Settable range: 2 minutes–30 minutes; factory
setting: 6 minutes.
Note: It is possible to register alphanumeric
characters (maximum 8 digits) as the Unique ID.
Refer to page 41.
Note: Relation between log interval and the estimated amount of data that can be saved:
Log interval=2 minutes: Approx. 46 days
Log interval=6 minutes: Approx. 135 days
Log interval=30 minutes: Approx. 664 days
When the amount of data saved exceeds the maximum, the oldest data is deleted and overwritten.
Log). Press
density, and open/closed state of
2
37

Publicidad

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

Mco-230aicuv

Tabla de contenido