INTRINSIC SAFETY

Intrinsically safe equipment is defined as "equipment and wiring which is incapable of releasing sufficient electrical
or thermal energy under normal or abnormal conditions to cause ignition of a specific hazardous atmospheric mixture
in its most easily ignited concentration." (ISA-RP12.6) This is achieved by limiting the amount of power available to
the electrical equipment in the hazardous area to a level below that which will ignite the gases.
In order to have a fire or explosion, fuel, oxygen and a source of ignition must be present. An intrinsically safe system
assumes the fuel and oxygen are present in the atmosphere, but the system is designed so the electrical energy or
thermal energy of a particular instrument loop can never be great enough to cause ignition.
BATTERIES
Batteries must be changed in Non-hazardous area.
Batteries must be of correct approved type.
Batteries must not be mixed with batteries of other models or manufacturers.
Batteries must be installed with correct polarity making sure the (+) end of the battery is aligned with (+)
symbol embossed in the battery case.
CERTIFIED Batteries for the TL3-A, TL3-R and TL3-W are as follows:
Model
TL3-A
TL3-W, TL3-R
Type
CR2032 LiMnO
2
AAA (LR03) Alkaline
Duracell
DL2032
MN2400
Panasonic
CR2032
LR03XWA
11/2020, JK