HOWEVER, THESE UNITS CAN DELIVER DANGEROUS LEVELS OF ELECTRICITY AND ARE TO BE OPERATED
ONLY BY QUALIFIED PERSONNEL FOLLOWING THE GUIDELINES LAID OUT IN THIS INSTRUCTION MANUAL.
ANYONE INTENDING TO USE THIS EQUIPMENT SHOULD READ THE COMPLETE MANUAL THOROUGHLY.
THE UNIT MUST NEVER BE USED WITHOUT THE SAFETY LID CORRECTLY IN POSITION.
THE UNIT SHOULD NOT BE USED IF THERE IS ANY SIGN OF DAMAGE TO THE EXTERNAL TANK OR LID.

Cleaning and maintenance

HORIZONTAL UNITS
-
Use warm water, but not above 60°C, as it can damage the unit and components
-
Use a mild detergent or water with a mild concentration of soap.
-
Do not leave the units in detergent for more than 30 min.
-
Compatible detergents include dishwashing liquid, Hexane and Aliphatic hydrocarbons.
-
Never let the following cleaning agents come into contact with the unit:
→ Acetone
→ Phenol
→ Chloroform
→ Carbon tetrachloride
RNase DECONTAMINATION
This can by following the instructions below:
-
Clean the units with mild detergent as described above
-
Wash with 3% hydrogen peroxide (H
-
Rinse with 0.1% DEPC (diethyl pyrocarbonate) treated with distilled water.

SAFETY PRECAUTIONS

WHEN USED CORRECTLY, THESE UNITS POSE NO HEALTH RISK.
CAUTION: DEPC is a suspected carcinogen. Always
take the necessary precautions when using it.
→ Methanol
→ Ethanol
→ Isopropyl alcohol
O
) for 10 min.
2
2