Troubleshooting; Replacement Parts - DENTAURUM Dublitherm compact Modo De Empleo

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

7

Troubleshooting

The following errors may occur:
Error
- Machine does not mix
- No LED display
- Appliance not mixing
- No LED lighting and no display in
the control pad.
- Appliance will not start
Error Message
CO
E-OL
E-SE
If you are unable to correct a fault on the basis of this guide, please contact your service technician or contact
Dentaurum Service direct:
8

Replacement Parts

Mains Cable
Fuse, 4.0 A TH
Teflon housing
Control PCB
Mixing paddle, Front
Mixing paddle, Rear
Knurled Nut for Mixing paddles
Cause
(1) appliance is not plugged in at
the mains
(2) appliance is not switched on
(3) faulty fuse
(4) Overheating, power supply cut
Lid not closed properly
Cause
Lid not closed properly
Motor overloaded
Temperature sensor fault
REF 907-027-10
REF 907-681-20
REF 971-503-00
REF 907-681-05
REF 907-696-01
REF 907-696-10
REF 907-696-60
Corrective Measures
(1) plug in at the mains
(2) switch the machine on
(3) change the fuse
(see Section 6.3!)
(4) allow the appliance to cool
down
Close lid
Orrective Measures
Close lid
Turn off appliance and allow to
cool down. Restart the appliance.
Contact Dentaurum Service Personell
dublitherm compact
en
EN
37

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

094-782-00

Tabla de contenido