Dc Current Measurement / A - Berner 313 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 55
-
Input resistance: 10 MΩ
-
Max. input voltage range of 200 mV to 250 V
-
Max. input voltage: 600 V AC RMS, frequency range:
40-400 Hz

DC current measurement / A=

Use the selector switch to set the appropriate range. Connect
the black test lead into the "COM" socket, and the red test
lead with the mA / 10A socket (the appropriate socket opens
automatically depending on the measuring range). Using the
test probes, touch the test points of the test object. Read
measurement value on the display. The current direction is in-
dicated by the sign.
You must use the '10 A' socket when measuring
currents above 200 mA.
Direct current
Measurement
range
2 mA
20 mA
200 mA
10 A*
Overload protection: mA range is protected with a 200 mA
self-resetting fuse.
-
10A range is protected with F 10 A / 1000 V.
-
In the 10A range observe the maximum operating times!
*To protect against overheating, measurements
should not be taken for more than 10 seconds.
Following this the device should be allowed to
cool down for at least 15 minutes.
Resolution
1 µA
0.01 mA
0.1 mA
0.01 A
Accuracy
± 1.0 % of reading + 3 digits
± 1.0 % of reading + 3 digits
± 1.5 % of reading + 5 digits
± 2.0 % of reading + 10 digits
30

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

185139

Tabla de contenido