OPERATION

CONTROL PANEL:

1. Main switch / Timer selector (optional).
2. Thermostat control knob.
3. Mains indicator lamp (green).
8
4a. Heater operation indicator lamp (amber) 50%.
4b. Heater operation indicator lamp (amber) 100%.
5. Timer knob (optional).
6. Safety thermostat control knob (optional).
7. Safety thermostat operation indicator lamp (optional).
8. Analog temperature reading thermometer.
9. Thermostat blockage device (optional).
11
10. Indicator lamp end of cycle.
11. Power selector 50% / 100%.
12. Safety thermostat reset push-button.
NOTE: The panel shown includes timer and safety thermostat,
5
supplied as accessories.
STARTING UP:
1. Press the mains switch (1). The green indicator lamp will
light up (3).
6
2. Turn the temperature control knob (2) clockwise until the
working temperature coincides with the pointer. The amber
4b
indicator lamp (4) will light up.
2
3. Block the temperature control knob (2) with the blockage
device (9) to avoid accidentally moving the knob.
4. After a certain time, the heater operation indicator lamp (4)
will start operating intermittently. When these intermittences
become uniform, the temperature of the oven is stabilized.
5. The thermostat scale is intended as a guide, the real tem-
perature level inside the oven is the one that the thermom-
eter (8) shows.
*
This oven is equipped with a power selector (11) which
permits the use of the oven at 50% or at 100% of its power.
ADJUSTING THE SAFETY THERMOSTAT:
To adjust the safety temperature thermostat, do as follows:
1. With the oven connected to the mains, turn the safety ther-
mostat control knob (6) clockwise to the end of the scale.
2. Ensure that the safety thermostat is reset by pressing the
reset button (12). The red indicator lamp (7) must be off.
3. Select the desired safety temperature with the temperature
control knob (2). This temperature has to be minimum at
5ºC above the working temperature.
Autovía A2 Km 585.1 Abrera 08630 (Barcelona) España