Potential For Radio/Television Interference; Troubleshooting - HoMedics THERA-P BPA-300 Manual

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

BPA-300-A.qxd
10/5/06
4:20 PM
POTENTIAL FOR RADIO/TELEVISION
INTERFERENCE
For USA only:
This product has been tested and found to comply with the limits for a
Class B digital device, pursuant to part 15 of FCC rules. These limits are
designed to provide reasonable protection against harmful interference in
a residential installation. The product generates, uses, and can radiate
radio frequency energy and, if not installed and used in accordance with
the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will not
occur in a particular installation. If the product does cause harmful
interference to radio or television reception, which can be determined by
turning the product on and off, the user is encouraged to try to correct
the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the product and the receiver.
• Connect the product into an outlet on a circuit different from that to
which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
For Canada only:
This digital apparatus does not exceed the Class B limits for radio noise
emissions from digital apparatus as set out in the interference-causing
equipment standard entitled "Digital Apparatus", ICES-003 of the
Canadian Department of Communications.
Cet appareil numérique respecte les limites de bruits radioeléctriques
applicables aux appareils numériques de Classe B prescrites dans
la norme sur le matériel brouilleur: "Appareils Numériques",
ICES-003 édictée par le ministre des communications.
Page 29
29

TROUBLESHOOTING

If any abnormality arises during use, please check the following points
POSSIBLE
SYMPTOMS
CAUSES
Unit does not turn on
Batteries have
when the
run down.
START/STOP
button is pushed.
Battery polarities
have been positioned
incorrectly.
EE measurement
Cuff has been placed
error symbol shown
incorrectly.
on display or the
blood pressure value
Did you talk or move
is displayed
during measurement?
excessively low
Shaking of the arm
(or high).
with the cuff on.
E1 error symbol
Air circuit abnormality.
shown on display
Cuff tube may not be
plugged into monitor
correctly.
E2 error symbol
Inflation pressure
shown on display
exceeding 300 mmHg.
E3 error symbol
Data error.
shown on display
CA appears on
Recalibration alert.
display
Note: If the unit still does not work, contact HoMedics Consumer
Relations. Under no circumstance should you disassemble or
attempt to repair the unit by yourself. Contact information for
HoMedics Consumer Relations Department can be found on
the warranty page.
30
CORRECTION
Replace them with
four new AA alkaline
batteries.
Re-insert the batteries
in the correct
positions.
Wrap the cuff
properly so that it is
positioned correctly.
Measure again.
Keep arm steady
during measurement.
Check cuff connection.
Measure again.
Turn the unit off, then
measure again.
Remove and reload
the batteries.
Refer to page 26 for
more information.

Publicidad

Capítulos

Tabla de contenido
loading

Este manual también es adecuado para:

Bpa-300

Tabla de contenido