BYK LCM IV Manual Básico Del Usuario página 22

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 103
Error displayed
No valid data found!
No measurement data present!
Control range not reached!
Control range exceeded!
Concentration too high!
Possible interference by:
Possible interference from:
Next service is due!
Negative result!
Network switched off.
Remote server cannot be reached.
Unstable lighting conditions!
Please insert adapter Z.
Insufficient memory for update .
System check incorrect!
Temperature too high.
Measurement not possible!
English 21
Definition
Data cannot be viewed in the data log
Data analysis settings cannot be configured without
measurement data.
Data analysis limits not reached
Data analysis limits exceeded.
Calculated concentration is higher than 999999
Interference Check
Interference Check
The calculated result is negative
Network setup off, when access to homepage via
sidebar
Error during network setup
For measurements with 11 mm round cuvettes, adapter
Z is required.
Error during update.
Measurement of air values failed
Solution
Change the selection.
Change the selection.
This is a warning notice. The control limit set was not
reached.
This is a warning notice. The control limit was
exceeded.
Dilute sample and measure again
The analysis is possibly erroneous due to
interferences.
The analysis is possibly erroneous due to
interferences.
Contact the manufacturer or national agent for an
instrument service.
Check concentration of sample
Activate the online connection.
Make sure that the instrument is connected to the
network.
Avoid direct sunlight at the measuring location.
Insert adapter Z into cell compartment (2). Confirm with
OK.
Select a memory with more space.
Switch the instrument off and then back on again. If the
system test is unsuccessful, contact the manufacturer
or the national agent.
Switch of the instrument and allow it to cool for a few
minutes. If necessary, move it to a cooler place.

Publicidad

Tabla de contenido
loading

Productos relacionados para BYK LCM IV

Este manual también es adecuado para:

Lcs iv

Tabla de contenido