7.2. Decontamination

7.2.1. Manual cleaning

E
• The contra-angle can be cleaned either by meticulous scrubbing under
running water, by steaming or by using wipes treated with microbiologically
tested detergent/disinfectant products (pH between 2.5 and 9).
• Regardless of the type of product used, observe the treatment time of the disinfectant
product used.
• Do not use alkaline or chlorine-based disinfectant products.
• Do not immerse the contra-angles in disinfecting baths or in an ultrasound basin.
• Do not scratch the surface of the optical fiber, as there is a risk that this may alter
the transmission of light.
7.2.2. Cleaning in machine
• Pre-wash with water.
• Wash with a substance at pH approximately 9 (weak alkaline), concentration 0.5% at
45°C for 5 minutes.
D
• Follow the thermal disinfector manufacturer's instructions for use, in
particular the recommendations regarding products wich may or may not be
used with the device.
• Do not immerse the contra-angles.
7.2.3 Cleaning the sprays
• In the event of an absence of liquid, the contra-angle must immediately be taken
out of use.
• Clean the spray nozzles carefully, using the metallic wire ref. no.
9205, in order to remove dirt and residue.
• If the nozzles are blocked, please contact the After-Sales Department.
r
WARNING : Risk of injury from the metal wire (9205). Wear protective gloves
and eyeprotection. If the nozzle remains blocked, please contact the After-
Sales Department or your approved repair workshop, or replace the nozzle.
In order to avoid any stagnation of the water in the shaft of the contra-angle, we
recommend that you blow air into the cooling liquid shaft. This is in order to avoid any
build-up of residue in the shafts. To do this, use one of the substances available to
disinfect spray channels. Wear eyeprotection and a face mask.
7.2.4 Cleaning the fiber optics
E
Do not scratch the surface of the fiber
optics, as there is a risk that this may alter
the transmission of light.