Battery Check - RS Pro RS-125 Manual De Instrucciones

Medidor de humedad o termohigrometro
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Moisture Meter/ English
Function check
"O------T------O"
Connect electrodes with T contacts to the protective cap.
Reference display for wood: 27%±2%
Reference display for building material: 1.25%±0.1%
Function error: Send instrument for servicing.
Application
• Remove the cap from the top side and place on the bottom side.
The instrument switches on automatically.
• Press the measuring electrodes as far as possible into the material.
• Always measure the moisture in the wood perpendicular to the fibre structure.
• Repeated measuring yields represent measured data.
• Once the measurement is finished, reattach the cap on the top side and the instrument
switches itself off.
Display
The reading is shown in the form of bars:
a continuous line
= even values e.g. 6, 8, 14
——dotted line
= uneven values e.g. 7,11
Overflow: reading >44%/2.0%
Please note: Characteristics due to growth or type of wood and deviating wood temperatures
may make it necessary to correct the displayed values. Measurements should be carried out
according to the drying and weighing/calcium carbide method in the case of inspections or for
reference purposes.
Automatic Power OFF
In order to conserve battery life, the meter will automatically turn off after approximately 15
minutes. To press and restoration the power switches the meter display is on.
2

Battery check

" O...... B......O"
Connect electrodes with B contacts to the protective cap.
Reference display for wood: >44%
Reference display for building material: >2.05%
Display has to full.
Changing the battery
When the batteries become exhausted or drop below the operating voltage, the battery
warning "
Remove the 1 screws at the back of the instrument.
Remove the battery housing. Change battery. Observe polarization! Reconnect instrument.
30/11/2017 Version No. 001
30/11/2017 Version No. 001
" symbol will appear in the LCD display. The battery should be replaced.
Moisture Meter/ English
3

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

146-9081

Tabla de contenido