VI. Changing the Unit of Measure
1. Press the
measure.
2. If the scale is turned off then back on, it will default to the
last unit used.
• ONYX-5K Units: g (gram) lb:oz (pound:ounce) oz
(ounce) ml (milliliter)
VII.
Calibration Adjustment (Advanced Users Only)
Calibration is performed at the factory. Only advanced users
should attempt calibration if the scale is not weighing properly.
1. Turn the scale on and wait for zero to be displayed.
2. Tap the
tap. A random number will be displayed.
3. With no weight on the scale, tap the
the zero point. After a couple of seconds the display will
show "ZERO" followed by "0000" (±2).
4. Place a 2000g calibration weight on the center of the
platform. After a couple of seconds the display will show
"CAL1" followed by another random number (ISN).
5. Now place a total of 5000g on the center of the scale
platform. The display will show "CAL2" then "5000g".
Calibration is complete!
key to change the scale's displayed unit of
fl.oz. (fluid ounce - US)
key four times, holding the key on the last
key again to set