Descargar Imprimir esta página

Faq - English; Corrective Action - Renfert Basic eco Manual De Instrucciones

Faq
Ocultar thumbs Ver también para Basic eco:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 7

FAQ - English

For information on operating elements, see
For chapter indications, refer to the complete instruction manual on the internet.
Problem
No air and no abrasive.
Only air, no or very
little abrasive.
Pressure control can-
not be adjusted.
Unit blasts continu-
ously.
Lamp is not alight.
Compressed air leaking
from the foot switch.
Foot switch fails to
switch on.
Foot switch hoses can-
not be disconnected
from the unit.
All manuals and user guides at all-guides.com
Possible cause
• Blasting nozzle blocked at the hand-
piece.
• Foot switch connecting lines kinked.
• Tank selection switch in an interme-
diate position.
• The dosage nozzle in the mixing
chamber is blocked due to contami-
nated or moist sand.
• Mixing chamber filter combination
blocked.
• Abrasive contains impurities or is
moist.
• The pressure control button is
locked.
• Foot switch connected incorrectly.
• Valve in foot switch is defective.
• Power cord not plugged in.
• Illumination defect.
• Pneumatic hoses reversed.
• No pressure at the compressed air
connection.
• Pneumatic hoses are kinked.
• Push-in connection is not released
(Fig. 2 /
3).
Fig. 1
and
Fig. 2
on page 2 of this document.
• Unscrew the nozzle from the hand-piece and blow it
out.
• With the nozzle removed, blow out the blasting hose
by activating the foot pedal.
• Inspect / correct the connection lines.
• Turn the switch until you feel it click into place.
• Cleaning the dosing nozzle, see chapter 6.5.
• Change filter cartridge
• Replace the abrasive.
• Pull the pressure control button upwards to unlock.
• Inspect / correct the foot switch connection.
• Replace foot switch.
• Plug the power cord into the wall outlet.
• Change light using the service or see spare part list.
• Reverse the connections.
• Inspect the compressed air lines and/or the compres-
sor.
• Reposition the hoses.
• Press the push-in connection ring on the hose side
firmly while pulling on the hoses.
- 1 -

Corrective action

(17, Fig.
1) (wrench SW14).
EN

Publicidad

loading