WARNING
ELECTROMAGNETIC FIELD INTERFERENCE
This instrument uses magnetic field method to measure the coating thickness on ferrous metal base.
If this meter was placed in the environment with 20mG (mini Gauss) or above, the accuracy would be
affected. Suggest that the meter should to put far away from the interfered source at least 30cm.
Elektromagnetic Source
Cellular Phone Charger
Notebook Power Supply
LCD-Display
Fan
Reading Lamp
Recommended operating conditions
1
SPECIFICATION
Display: 3½ digit liquid crystal display (LCD) wit maximum reading of 1999.
Low Battery Indication: The " " is displayed when the battery voltage drops below the operating level.
Measurement Rate: 1 second, nominal.
Operating Environment: 32°F to 122°F (0°C to 50°C) at <75% R.H.
Storage Temperature: -4°F to 140°F (-20°C to 60°C),0 to 80% R.H. with battery removed from meter.
Auto Power Off: 30 seconds.
Standby Consuming Current: < 15µA.
Battery: Standard 9V battery (NEDA 1604, IEC 6F22 006P).
Battery Life: 9 hours (continuity) typical (contain Backlit).
Dimensions: 148 mm (H) x 105 mm (W) x 42 mm (D).
Weight: Approx. 157g (including battery).
Detectable Substrate Material: Ferrous metal (iron, steel) and Non-Ferrous metal (copper, aluminum,
zinc, bronze, brass, etc.)
ELECTRICAL
Thickness Range: 0 to 40.0mils (0 to 1000µm).
Display Resolution: 0.1mils/1µm.
Accuracy:
±4dgts on 0 to 7.8mils
±10dgts on 0 to 199µm
±(3%+4dgts) on 7.9mils to 40mils
±(3%+10dgts) on 200µm to 1000µm
Temperature Coefficient: ±0.1% of reading, whichever is greater, change in accuracy per °F/°C
change in ambient operating temperature above 82.4°F/28°C or below 64.4°F/18°C.
Response Time: 1 second.
BGS technic KG
Bandwirkerstr. 3
D-42929 Wermelskirchen
Electromagnetic field strength (unit = mini Gauss)
Any product with coil inside should be considered.
( >30cm)
210 m
µ
A L M
µm
F e r r
A
H
M A X
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
22
23
24
25
26
27
28
29
30
mG
Tel.: 02196 720480
Fax.: 02196 7204820
mail@bgs-technic.de
0 cm
50 ~ 500
100 ~ 1000
10 ~ 100
100 ~ 1000
400 ~ 4000
Abnormal operating conditions (<30cm)
30 cm
< 1
< 5
< 5
< 5
< 10
210 m
µ
A L M
µm
F e r r
A
H
M A X
mG
www.bgstechnic.com