Important note for calibration (deleting the calibration data)
If the calibration fails, first perform the following steps.
Then re-perform all the steps from 2-point calibration.
Push and hold "Zero" button.
1.
Pull trigger
2.
Display shows „Clr" und "Set".
Calibration data be erase.
3.
Start a new 2 point calibration.
Frequent Calibrating Point
Push and hold "MAX/MIN" button.
1.
Pull trigger
2.
Display shows "SET" und "dFut ".
Push „▲" or „▼" button to set calibration value.
3.
example: 39,6 mils (1006 µm)
"CAL" Taste drücken
4.
Calibration value is stored.
Put standard coating plate on measuring plate.
5.
(Standard coating plate 39,6 mils / 1006 micron)
Push sensor on standard coating plate and measuring
plate. Pull trigger.
6.
(Take the ferr or non-ferr measuring plate for calibrate)
Wait until the display shows the
7.
Push „CAL" button