• Do not use bent, damaged, or sub-standard file or reamer. The shank could bent or broke even within the recommended speed.
• Always keep the file, reamer shank clean, Entry of debris into the chuck cause file or reamer run-out and the file or reamer could not be mounted inmost.
• DO NOT exceed the maximum permissible rotation speed that manufactor recomend.
• Connect micro motor for X35 series and X75 series, Air motor for X70 series.
• Do not use the file, reamer or shake shape other than ISO 1797-1 Type1 standards.(Except Hand file)
• File may change (throwaway) every patient or several times. If it used several time, it may cause a fracture by metal fatigue.
FOR X57L,X57
• Wash the handpiece immediately after use. If you do not perform it, abrasive paste may fix inside and it may cause a failure.
• Check the ring before you use. If the ring loosen, tighten it by the wrench.
• When mounting the rubber cup, check it correctly attached.
• DO NOT use the rubber cap which the screw does not match.
• Do not use above the specified rotational speed. This could cause in injury or accident by flying paste or broken cup and brush.
• Do not use in the reverse rotation. The cup and brush could fly out of the head