5. Technical Data
ErgoOne
®
Single-Channel
0.1 – 2.5 µl
0.5 – 10 µl
2 – 20 µl (micro cone)
2 – 20 µl (standard cone)
10 – 100 µl
20 – 200 µl
100 – 1,000 µl
500 – 5,000 µl
ErgoOne
®
Multi-Channel
0.5 – 10 µl
10 – 100 µl
30 – 300 µl
Measurement data determined in compliance with the test conditions for piston-stroke pipettes
according to EN ISO 8655
Measurement carried out with original TipOne
12
ErgoOne
Technical Data ErgoOne
Systematic measurement
deviation
Pipette
(inaccuracy) ± [%]
volume [µl]
EN ISO
8655-2
0.1
± 125
0.25
± 50
1.25
± 10
2.5
± 5.0
1.0
± 12
5
± 2.4
10
± 1.2
2
± 10
10
± 2.0
20
± 1.0
2
± 10
10
± 2.0
20
± 1.0
10
± 8.0
50
± 1.6
100
± 0.8
20
± 8.0
100
± 1.6
200
± 0.8
100
± 8.0
500
± 1.6
1,000
± 0.8
500
± 8.0
2,500
± 1.6
5,000
± 0.8
Systematic measurement
deviation
Pipette
(inaccuracy) ± [%]
volume [µl]
EN ISO
8655-2
1.0
± 24
5
± 4.8
10
± 2.4
10
± 16
50
± 3.2
100
± 1.6
30
± 26.7
150
± 5.3
300
± 2.7
pipette tips
®
User Manual
®
Pipettes
®
Coincidental measure-
ment deviation
(imprecision; CV) ± [%]
EN ISO
ErgoOne
®
8655-2
± 24
± 75
± 12
± 30
± 2.5
± 6.0
± 1.4
± 3.0
± 2.5
± 8.0
± 1.5
± 1.6
± 1.0
± 0.8
± 5.0
± 5,0
± 1.2
± 1,0
± 1.0
± 0,5
± 5.0
± 5,0
± 1.2
± 1,0
± 1.0
± 0,5
± 3.0
± 3.0
± 1.0
± 0.6
± 0.8
± 0.3
± 2.5
± 3.0
± 1.0
± 0.6
± 0.6
± 0.3
± 3.0
± 3.0
± 1.0
± 0.6
± 0.6
± 0.3
± 2.4
± 3.0
± 1.2
± 0.6
± 0.6
± 0.3
Coincidental measure-
ment deviation
(imprecision; CV) ± [%]
EN ISO
ErgoOne
®
8655-2
± 8.0
± 16
± 4.0
± 3.2
± 2.0
± 1.6
± 3.0
± 6.0
± 1.0
± 1.2
± 0.8
± 0.6
± 3.0
± 10
± 1.0
± 2.0
± 0.6
± 1.0
ErgoOne
®
± 12
± 6.0
± 1.5
± 0.7
± 1.8
± 0.8
± 0.4
± 1.5
± 0.6
± 0.3
± 1.5
± 0.6
± 0.3
± 1.0
± 0.3
± 0.2
± 0.7
± 0.3
± 0.2
± 0.6
± 0.2
± 0.2
± 0.6
± 0.25
± 0.15
ErgoOne
®
± 5.0
± 2.0
± 1.0
± 2.0
± 0.8
± 0.3
± 1.0
± 0.5
± 0.3