Please read and follow these instructions for use and keep
them for future reference.
Intended use
The HEINE OMEGA 600 and HEINE OMEGA 600 wired binocular
indirect ophthalmoscope is a head worn AC-powered or battery
powered device for transient use, containing illumination and viewing
optics intended to examine posterior segments of the eyes.
The product must only be used by qualified medical professionals
and in medical healthcare facilities.
For U.S. only:
Federal law restricts this device to sale by or on the order
of a Physician or Practitioner.
Warnings and safety information
CAUTION! This symbol indicates potential hazardous situations.
Ignoring the corresponding instructions may lead to dangerous
situations of mild to moderate extent. (Background color yellow;
foreground color black).
NOTE! This symbol indicates valuable advice. Notes are
important, but not related to hazardous situations.
Product overview
OMEGA 600
Fig. 1
3
1
10
7
9
8
OMEGA 600 wired
Fig. 2
3
12
11
4
Optics unit for OMEGA 600 and OMEGA 600 wired
Fig. 3
18
15
17
16
Rechargeable battery CB1
Fig. 4
6/20
2
3
3
6
4
2
3
3
5
6
14
15
19
17
19
10
1 Width adjustment
2 Height adjustment
3 Padding
4 Socket for brightness control
5 Adjustment lever
6 Brightness control
7 Charge status indicator
8 Battery compartment
9 USB-C socket
10 Rechargeable battery CB1
11 Power indicator
12 Power supply connection
13 Optics unit
14 Aperture selection lever
15 Illumination height adjustment
16 Dust cover
17 Stereoscopic adjustment lever
18 Filter selection lever
19 Eyepieces
Setting up
To operate the OMEGA 600, insert the rechargeable battery CB1 (10)
into the battery compartment (8) of the instrument until the battery
CB1 (10) snaps into place. We recommend to charge the battery CB1
(10) of the device fully before first use.
Remove the insulation film between the battery CB1 (10) and the
battery compartment (8) of the device.
To operate the OMEGA 600 wired, connect the power supply unit with
an appropriate plug into a mains power socket.
The device must only be operated at the mains voltage specified
on the type plate.
Mount the power supply in such a position where it can be easily
5
unplugged. The LED indicates the operational readiness of the power
supply. The power indicator (11) is automatically activated when the
device is switched on.
4
To put the power supply out of operation just disconnect the power
supply from the mains supply.
Do not pull the cable to disconnect the power adapter from the
mains.
Charging of the OMEGA 600
Charge the device outside of the patient environment (at least
1.5 metres from the patient or patient support pursuant to IEC
60601-1, see Fig. 5).
Fig. 5
Charging via USB
For charging, connect the USB power supply unit (e.g. HEINE E4-
USB) to the USB-C socket (9) of the device.
In charging mode, the LED lighting is switched off and the illumi-
18
nation of the device is no longer supplied with power.
Do not let the power cable become taut as this could damage
the device or pose a tripping hazard.
The setting up and operation of the HEINE E4-USB is described in a
separate instruction for use.
Charging via wall charger CW1
Use the wall charger CW1 to charge the battery CB1 (10) of the
device. The blue LED on the wall charger CW1 indicates contact
with the device. The charge status indicator (7) is located on the
15
battery compartment (8) of the device.
Make sure that no foreign objects come between the device and
the wall charger CW1. When using the wall charger CW1 to charge
the device, make sure that the wall charger CW1 is not contami-
nated in any form.
The setting up and operation of the wall charger CW1 is described in
a separate instruction for use.
MED 234766 2020-12-15
ENGLISH