Cleaning Instructions; Suction Unit; Carrying Case - DeVilbiss Vacu-Aide 7314 Serie Guía De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 8
Filter Cartridge Changing
1
2
Remove filter cartridge and
Turn unit
"off".
43/8" tubing.
CLEANING INSTRUCTIONS
Collection Container
1
2
Disconnect from power source. Disconnect tubing and remove
Turn unit
"off" and allow
vacuum to drop.
NOTE– If float ball was
7
removed for cleaning,
ensure it is properly
installed in lid prior to use.
Wash in 1 part vinegar to 3
parts warm water solution.
Rinse with clean, warm water
and air dry.
6' Patient Tubing
1
2
Disconnect from lid.
Rinse thoroughly by running
warm tap water through it.
Suction Unit
1
2
Disconnect from power source. Wipe the housing with a clean
Turn unit
"off" and allow
vacuum to drop.
CAUTION– Do not submerge in water as this will result in damage to the vacuum pump.
A-7314
DANGER
3
Electric shock hazard. Do not attempt to open or remove cabinet, there are no
user-serviceable internal components. If service is required, return unit to a
qualified DeVilbiss healthcare provider or an authorized service center. Opening
or tampering with the unit will void warranty.
NOTE– Do not substitute any other material for this filter cartridge. Substitution may
lead to contamination or poor performance; use only DeVilbiss filter cartridges.
NOTE– The filter cartridge contains a hydrophobic filter. If the filter media becomes wet,
air flow will be stopped. The filter cartridge must then be replaced. Do not remove filter
Install new cartridge and
media from filter cartridge.
attach tubing.
3
4
Carefully remove lid and
container from holder.
empty contents. NOTE– The
800ml disposable collection
container and lid are meant
for single-patient use.
3
4
Follow by running a solution of
Keep outer surface clean by
1 part vinegar to 3 parts warm
wiping with clean, damp cloth .
water through it. Rinse with
warm tap water and air dry.
3
cloth and any commercial
(bacterial-germicidal)
disinfectant.
5
6
Remove filter cartridge and
Wash container and lid in
warm water/dishwashing
43/8" tubing and set aside.
solution. Rinse with clean,
warm water. NOTE– If desired,
the float ball can be removed
for washing by pushing ball
out of lid assembly.
AC to DC Adapter
1
2
Disconnect AC to DC adapter
Wipe AC to DC adapter
from unit and from power
housing and cords with a dry
source.
cloth.
Carrying Case
1
Wipe with clean cloth
dampened with detergent or
disinfectant.
EN - 5

Publicidad

Tabla de contenido
loading

Tabla de contenido