ELECTROMAGNETIC COMPATIBILITY
Guidance and manufacturer's declaration - electromagnetic immunity
Emissions Test
Electrostatic discharge
(ESD) IEC 61000-4-2
Electrical fast transient/
burst IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short in-
terruptions and voltage
variations on power
supply input lines IEC
61000-4-11
Power frequency
magnetic field
IEC 61000-4-8
Conduced RF
IEC61000-4-6
Radiated RF
IEC61000-4-3
NOTE U T is the a.c. mians voltage prior to application of the test level.
20 • ENGLISH
Table 2 EMISSIONS
IEC 60601-1-2
test level
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Not application
Not application
Not application
30 A/m
50Hz/60Hz
Not application
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Compliance
level
±8 kV contact; ±2 kV, ±4
kV, ±8 kV, ±15 kV air
Not application
Not application
Not application
30 A/m
50Hz/60Hz
Not application
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz