5.4 Environmental Specifications
Temperature
Operating
Storage/Transportation
Humidity
Operating
Storage/Transportation
Atmosphere Pressure
Operating
Storage/Transportation
5.5 Physical Specifications
Width x Height x Depth
Weight
5.6 Display
Display type
Display content
5.7 LED Wavelength
RED
IR
6. Technical Description
The below table shows statistic conclusion of an invasive controlled desaturation study, which guided by
"ISO80601-2-61, Annex EE, Guideline for evaluating and documenting SpO2 Accuracy in human
subjects". The statistic result displayed the accuracy distribution between the range of 70% ~ 100%,
which may be helpful to user.
Bias Analyse
SpO2-FS20P
Pulse Oximeter
Mean Bias (B
Precision (S
Accuracy (A
The below is the Bland-Altman graphical plot of samples from invasive controlled desaturation study.
7. Maintenance
7.1 Maintenance
The equipments design life expectancy is about 2 years, keep your equipment and accessories free of
dust and dirt, and follow these rules :
A. Please clean the equipment before use according to chapter 8. Remove the batteries inside the
battery cassette if the equipment will not be operated for a long time.
Wavelength
660 ± 6nm
905 ± 10nm
SaO2-Radiometer ABL800 FLEX-CO-Oximeter
70 – 80 (%)
)
1.94
s
)
2.00
res
)
1.98
rms
Bland Altman Graph for SaO2-SpO2
+41°F - +104°F/+5°C - +40°C
-4°F - +140°F/-20°C - +60°C
10-95% non condensing
10-95% non condensing
70-106kpa
50-107.4kpa
About 53x29x30mm
About 21g (including batteries)
OLED 0.96'', 128x64 pixel
SpO2%, Pulse rate, battery indicator, bar graph,
Pulse waveform, heart beat mark
Radiant Power
1.8mW
2.0mW
80 – 90 (%)
90 – 100 (%)
1.45
0.89
1.55
0.98
1.53
0.96
EN
70 – 100 (%)
1.40
1.53
1.52