To discharge the battery pack completely, continue to run the
oscilloscope on the battery pack until automatic shutdown
occurs.
NOTE.
To prolong the life of the battery pack and to prevent
shutdown, do not operate or charge the battery pack at high
temperatures. For best results, allow the battery pack to cool to
room temperature before using or charging the battery pack.
Cold temperatures adversely affect the normal electrochemical
reactions within a battery pack, reducing the available capacity.
You can minimize this loss by keeping the battery packs at room
temperature prior to and during use.
Do not leave a battery pack discharged for extended periods.
See the Storing Battery Packs section for information on how to
properly store a battery pack.
The battery pack stops charging under the following conditions:
The battery pack is fully charged.
The battery pack temperature exceeds the operating
temperature specification in the table below.
The battery pack shuts down and disconnects itself from the
TDS3000, TDS3000B, and TDS3000C series oscilloscope
under the following conditions:
The battery pack is fully discharged.
There is a fault current that exceeds 5 A for longer than
32 ms.
Calibrating the Battery Gauge
It is recommended that you perform this procedure:
For all new batteries before using the battery.
On the TDS3BATC every 45 days during its useful service
life.
To calibrate the battery gauge, follow these steps:
1.
Completely discharge the battery through the oscilloscope.
2.
Charge the battery fully the first time either on the
oscilloscope (TDS3000, TDS3000B, or TDS3000C series)
for 32 hours or on the external charger (TDS3CHG) for 6
hours. If you use the external charger, then you can charge
the battery until the luminous indicator on the external
charger turns off.
3.
Operate the oscilloscope with the battery until the battery
is completely discharged (the oscilloscope turns off).
Copyright © Tektronix, Inc. All rights reserved.
NOTE.
Do not connect AC power to the oscilloscope during
this time.
The battery gauge is now calibrated and should accurately
indicate the level of charge in the TDS3BATC.

Replaceable Parts

Part
Part Number
Grounding wire
174-3997-00
Specifications
Characteristic
Description
Input from TDS3CHG
30 W, 18 VDC
Output
14.4 VDC, 6450 mAh
Operating Temperature
0 °C to +50 °C (+32 °F to +122 °F),
discharging
0 °C to +45 °C (+32 °F to +113 °F),
charging
Storage Temperature
-20 °C to +40 °C (-4 °F to +104 °F)
Handling Battery Packs
Do not disassemble, crush, or puncture a battery.
Do not short the external contacts on a battery.
Do not dispose of a battery in fire or water.
Do not expose a battery to temperatures above +60 °C
(+140 °F).
Keep the battery away from children.
Avoid exposing the battery to excessive shock or vibration.
Do not use a damaged battery.
If a battery pack has leaking fluids, do not touch any
fluids. Dispose of a leaking battery pack. See the Battery
Recycling section for information on Disposal and
Recycling.
In case of eye contact with fluid, do not rub eyes.
Immediately flush eyes thoroughly with water for at least
15 minutes, lifting upper and lower lids, until no evidence
of the fluid remains. Seek medical attention.
Storing Battery Packs
Store battery packs in a low-humidity environment free of
corrosive gases. Storing battery packs in high-humidity
environments, or outside the temperature range, can cause
oxidation on the metallic parts and increased leakage.
Charge or discharge the battery to approximately 50% of
capacity before storage.
Charge the battery to approximately 50% of capacity at
least once every six months.
Remove the battery and store it separately from the
oscilloscope.
For optimal battery life, store the battery at temperatures
between +5 °C and +20 °C (+41 °F and +68 °F).
NOTE.
The battery self-discharges during storage. Higher
temperatures (above +20 °C or +68 °F) reduce the battery
storage life.
Replacing Battery Packs
Use the instructions to remove and replace the battery packs.
NOTE.
Replace the Li-Ion battery packs only with TDS3BATC
battery packs.
See the Battery Recycling section for information on how to
properly dispose of a Li-Ion battery pack.

Warranty Information

For warranty information, go to the www.tektronix.com Web
site, and search for "TDS3BATC warranty."
Contacting Tektronix
Tektronix, Inc.
14200 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA
For product information, sales, service, and technical support:
In North America, call 1–800–833–9200.
Worldwide, visit www.tektronix.com to find contacts in
your area.
www.tektronix.com

Publicidad

Tabla de contenido
loading

Tabla de contenido