Installation - Selecta UV 3100 Manual De Instrucciones

Espectrofotómetro
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

INSTRUCTION MANUAL CODE 80260 REV A
J
.P. SELECTA s.a.u.
Feb/2017

4.3 Installation

Step 1: Check the packing list
Unpack the contents, check the materials with the Packing List. Any damage or so-
mething missing, please contact the local dealer or J.P.Selecta, s.a.
Step 2: Position
Place the instrument on a stable surface carefully.
Step 3: Install printer (optional)
Make sure the printer's power is off. Link the printer's data cable to the equipment's
parallel port.
Step 4: Connect the power cord
Make sure the instrument's power switch is in the Off condition, connect the power
cord to the equipment. And insert another end in the socket provided with a protec-
tive earth contact.
Step 5: Switch on the power
Check again. Make sure that all the cables are right. Switch on the power. Then the
instrument begins to self-test. After the self-test and 15 minutes pre-warm, it can
work. The self-test includes the following steps:
D2 Lamp Pre-warm --> Com-port Initialization --> Initialize Printer --> Start kernel
--> Initialize AD --> Filter position --> Pre-warm.
After pre-warming, the instrument will ask the user to re-calibrate the system. Users
can decide if they need to re-calibrate the system or not. After this step, the equip-
ment will work normally.
After transportation or a period of use, the working condition of the equi-
pment may changed because of environment changing or any other cause
(attenuation of lights' energy, vibration, etc.) and then users need to re-
calibrate the system to keep it in the best working conditions.
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
(It can be modified without notice)
Tel 34 93 770 08 77 Fax 34 93 770 23 62
Page: 38

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

4120021

Tabla de contenido